MEASUREMENT

87

Program

Overview

Background

The Program function measures input signals
using cursors or automatic measurement
functions, with a user-defined sequence, duration,
loop count, and panel settings. This feature is
useful for automated and repetitive measurement,
such as in assembly lines or quality inspection
tests.

Parameter

Program set

1 set

Program step

Maximum 20 steps

Measurement
item

Cursor or Automatic
measurement

Time (duration)
per step

1 ~ 99 seconds, or user activation

Program loop

1 ~ 99 loops, the first and last step
settable

Programming
step

1. Show the target waveform on the display and

decide the type of measurement that needs to
be done: Horizontal/Vertical Cursor or
Automatic measurement.

2. Setup the other panel configurations: trigger,

acquisition, horizontal/vertical scale, etc. Save
the settings to the internal memory. See
page146 for details.

3. Edit the program (page88) using the internally

stored panel setup.

4. Run the program (page90).