COMMAND SET

29

:CHANnel:SCALe

Set

Query

Description

Sets or returns the vertical scale for each channel. The
scale is dependent on the probe attenuation factor.
Same as: Volts/Div knob

Syntax

< Long >

< Short >

:channel:scale
:channel:scale?

:chan:scal
:chan:scal?

Parameter

Channel

Vertical scale

1/2/3/4 CH1/2/3/4 2e–3 ~ 5e+0 2mV ~ 5V

(Probe x1)

2e–2 ~ 5e+1 20mV ~ 50V

(Probe x10)

2e–1 ~ 5e+2 200mV ~ 500V

(Probe x100)

Example

:channel1:probe 0
:channel1:scale 2.00e–3

Set CH1 probe to x1
Set CH1 scale at 2mV