Additel 221A Multifunction Temperature Calibrator User Manual User Manual
Page 83

________________________________________________________________________________________________________________
76
trim, then you can press Enter or Done to display the Output Trim
interface (as shown in Figure 11.12).
3. In the Output Trim interface, press Enteror Trim to execute the
output point adjustment. The trim value can be fetched by pressing
Fetch, or it can be inputted manually.
4. If Output Trim is successfully executed, the calibrator will goes back
to the Output Trim point selection interface. You can repeat steps 2
and 3 to start another Output Trim.
Figure 11.12 HART Output Trim