Additel 221A Multifunction Temperature Calibrator User Manual User Manual
Page 57

________________________________________________________________________________________________________________
50
interpolation equations.
The calibrator uses the first and last deviation function for the calculation for SPRT.If you use other temperature ranges,
and
can be assigned to zero (i = 7, 8, 9, 10, 11).
The table 9‐1 list the sub ranges, required calibrated in the NIST SPRT Laboratory.
Table 9‐1
Temperature range(
℃)
Required Fixed Points
Deviation Function
-189.3442 to 0.01
Ar TP, Hg TP, TPW
∆
0 to 29.7646
TPW, Ga TP
∆
1
0 to 156.5985
TPW, GaTP,In FP
∆
1
0 to 231.928
TPW, In FP, Sn FP
∆
1
9
1
2
0 to 419.527
TPW, Sn FP, Zn FP
∆
1
8
1
2
0 to 660.323
TPW, Sn FP, Zn FP, Al FP
∆
On the RTDs Management page, you can delete an existing RTD by pressing Deleteand re‐edit it by pressing Enter or
Edit.
When the custom RTD is added to the RTDs Library successfully, your calibrator is now ready to make measurements
using your custom RTD.