Model
Features
761-LLP
761-D
761-L
761-LA
761-M
761-MA
761-H
761-HA
761-BP
Pressure
Range
-25 to 25 mbar
(-10 to 10 inH
2
O)
-0.95 to 1 bar
(-13.5 to 15 psi)
-0.95 to 7 bar
(-13.5 to 100 psi)
0.05 to 8 bar.a
(1.2 to 115 psi.a)
-0.90 to 25 bar
(-13 to 375 psi)
0.1 to 26 bar.a
(1.7 to 390 psi.a)
-0.90 to 40 bar
(-13 to 600 psi)
0.1 to 41 bar.a
(1.7 to 615 psi.a)
100 to
1200 hPa
Pressure
Module 1
-2.5 to 2.5 mbar
(-1 to 1 inH
2
O)
-25 to 25 mbar
(-10 to 10 inH
2
O)
-0.95 to 2.5 bar
(-13.5 to 35 psi)
0.05 to 3.5 bar.a
(1.2 to 50 psi.a)
-0.9 to 2.5 bar
(-13 to 35 psi)
0.1 to 3.5 bar.a
(1.7 to 50 psi.a)
-0.9 to 2.5 bar
(-13 to 35 psi)
0.1 to 3.5 bar.a
(1.7 to 50 psi.a)
100 to
1200 hPa
Pressure
Module 2
-25 to 25 mbar
(-10 to 10 inH
2
O)
-0.95 to 1 bar
(-13.5 to 15 psi)
0 to 7 bar
(0 to 100 psi)
0.05 to 8 bar.a
(1.2 to 115 psi.a)
0 to 25 bar
(0 to 375 psi)
0.1 to 26 bar.a
(1.7 to 390 psi.a)
0 to 40 bar
(0 to 600 psi)
0.1 to 41 bar.a
(1.7 to 615 psi.a)
NA
Differential
Pressure
•
•
Gauge Pressure
•
•
•
•
•
•
•
•
Absolute
Pressure
•
•
•
•
Barometric
Pressure
•
•
•
•
Temperature
Compensated
•
•
•
•
•
•
•
•
•
Built-in Electrial
Pump
•
•
•
•
•
•
•
•
•
Built-in Filter
•
•
•
•
•
•
•
•
•
Built-in Liquid
Trap
•
•
•
•
•
•
•
•
•
Source 0 to
22mA
•
•
•
•
•
•
•
•
•
Measure mA
or V
•
•
•
•
•
•
•
•
•
Measure
external
pressure
module
•
•
•
•
•
•
•
•
•
24V Loop power
•
•
•
•
•
•
•
•
•
Pressure switch
test
•
•
•
•
•
•
•
•
•
HART
Communication
•
•
•
•
•
•
•
•
•
Documenting
•
•
•
•
•
•
•
•
•
Automated Pressure Calibrators
Selection Guide
Additel 761