13
Simulating Thermocouple Source&using Thermocouple to Measure Temperature Accuracy. (Working temperature:
20℃±5℃)
Table5‐3
Thermocouple Measurement and Source accuracy
Measure and
Simulate
Standard
Temperature Range (
℃)
Accuracy (
℃)
Measure
Source
S
IEC 584
‐50 to 1768
‐50 to 400
1.0
1.1
400 to 1000
0.6
0.6
1000 to 1768
0.7
0.8
R
IEC 584
‐50 to 1768
‐50 to 200
1.4
1.4
200 to 500
0.6
0.6
500 to 1768
0.6
0.7
B
IEC 584
0 to 1820
50 to 450
3.8
3.8
450 to 800
0.9
0.9
800 to 1820
0.7
0.7
K
IEC 584
‐270 to 1372
‐250 to ‐200
1.0
1.1
‐200 to ‐100
0.4
0.5
‐100 to 600
0.3
0.3
600 to 1372
0.4
0.5