0 setpoints – Rice Lake Condec UMC600 User Manual
Page 58

Setpoints
55
7.0
Setpoints
The UMC600 is equipped with four independent control outputs, which are 5VDC logic level signals capable of
sinking 12mA. Setpoints can be configured in the following modes:
•
Mode A – Setpoint and zero band
•
Mode B – Setpoint, with preact, and zero band
•
Mode C – Setpoint, with dribble control
•
Mode D – Setpoint, with preact and dribble control
•
Mode E – One setpoint with preact, dribble control, and tolerance band output
•
Mode F – High-low alarm control
•
Mode G – Over/under checkweigh control
The following table describes the various modes.
Mode
Description
Setpoint
Mode A
Setpoint and zero band
Cutoff occurs when WT ≥ SP.
Use the following steps to set Mode A setpoint values:
1.
Press
SETPOINT 1
or
SETPOINT 2
on front panel.
2.
Enter desired setpoint value, (display times out if no number is entered for several
seconds).
3.
Press
ENT
, then display returns to normal operation.
4.
The setpoint value can be recalled any time by pressing the appropriate setpoint key.
SP1
SP2
ZBD
Mode B
Setpoint, with preact and zero band
The preact value provides for material that is in suspension. Cutoff occurs when WT ≥ SP – Pr
Use the following steps to set Mode B setpoint values.
1.
Press
SETPOINT 1
or
SETPOINT 2
on front panel.
2.
Enter desired setpoint value, (display times out if no number is entered for several
seconds).
3.
Press
ENT
. Display flashes number value alternated with Pr.
4.
Enter desired preact value, then press
ENT
key. Display returns to normal operation.
5.
The setpoint value can be recalled any time by pressing the appropriate setpoint key.
SP1-Pr
SP2-Pr
ZBD
100
90
80
70
60
50
40
30
20
10
0
ZBD
SP1
SP2
WEIGHT
MODE A
WIRE
PLACEMENT
TB3 - 5
TB3 - 4
TB3 - 3
TB3 - 2
100
90
80
70
60
50
40
30
20
10
0
ZBD
SP1-Pr
SP2-Pr
WEIGHT
MODE B
WIRE
PLACEMENT
TB3 - 5
TB3 - 4
TB3 - 3
TB3 - 2