3 digital configuration parameters – Rice Lake Condec UMC600 User Manual
Page 17

14
UMC600 Installation Manual
3.1.3
Digital Configuration Parameters
Table 3-1 (shown below) shows an overview of the 14 configuration parameters and their description.
Table 3-2 on page 16 lists the configuration display prompts (Prompt 1), and their value selections for displayed
graduations. Prior to calibration, the UMC600 must be digitally configured, or given its set of operating parameters.
The first three parameter selections are directly related to calibration and must be set up before proceeding to the
calibration mode. These parameters include;
•
number of graduation
•
resolution
•
decimal point location in the weight data, all of which define the scale capacity.
Parameter
Description
Values
1
Graduations
NTEP to 10,000 (up to 80,000 available).
2
Display resolution
1, 2, 5, 10, 20, 50, 100
3
Decimal point
0.0, 0.00, 0.000, 0.0000, no decimal
5
Digital averaging
1, 2, 4, 8, 16, 32, A1 = 8-4-2; A2 = 16-8-4
6
Tare mode
ATNR, AUTO, FIXED, BOTH (inhibit with motion)
7
AZM band
Off, 0.5, 1, 3, 5, and 10 divisions
Use 0.5 for H-44, bench, counter and livestock applications; use 3 for vehicle,
axle-load, and railroad scales
8
AZM/PAZ aperture
±1.9%, 100% of capacity includes push-to-zero, H-44: 1.9%
9
Motion
Off, 1, 3, 5 divisions H-44: vehicle, axle, livestock, RR, 3.0; all other 1.0
10
Displayed units
Lb, kg, con
11
Setpoint mode
See Section 7.0 on page 55 for settings
12
Zero band
Off, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 20, 30, 40, 50
13
Weigh mode
Normal, truck mode with transaction cancelled/stored, fixed/auto tare
14
Serial output
Port 1, Port 2, demand, continuous, baud rate, G/T/N, or display
Table 3-1. Parameter Overview