6 option 6 – linear calibration – Rice Lake Condec UMC600 User Manual
Page 35

32
UMC600 Installation Manual
ID HELD
This function is enabled when 5.3 is selected On. When ID held is enabled, the current keyed in ID number is stored
in memory and is sent out with each valid demand serial print. There are two methods of ID entry.
•
Gross/Net mode – use numeric keys to enter the desired ID number, then press
.
•
Gross mode only – key in numeric value and press
ENTER
. This stores the new ID number for the next
print command.
NOTE: Leading zero entry not allowed.
NOTE: If a zero value is entered as an ID number, there is no printout of an ID number.
Counter
The counter feature is a six digit counter that sequentially increments by one on each valid print command in the net
mode only. It is enabled when 5.4 is on and has a range from 0 - 999,999.
4.6
Option 6 – Linear Calibration
There are five calibration points that can be entered sequentially. Calibrations using fewer than five points will
linearize the curve up to the last data point.
Enabling option 6 places the indicator in a linearization calibration when the unit is in the calibration mode (SW1-1
closed on front panel DIP switch). Use the five-point linearization procedure outlined in Section 5.3 on page 48 to
perform a scale calibration when option 6 is enabled.
The five-point linearization calibration (OP.6) is shown in Table 4-9
.
Table 4-9. Option 6 Configuration Selections
Refer to Section 4.12.3 on page 46 for individual serial command option parameters for Option 6.
Prompt
Display
Interpretation
Notes
Five-point linearization
OP.6
oFF
Off
—
OP.6
on
On
Enter up to five linearization points