beautypg.com

Yokogawa Value Series FX1000 User Manual

Page 65

background image

2-52

IM 04L21B01-64EN

• Start channel and End channel (FX channels)

Enter the first and last channel numbers of input/output. The channel numbers that

you can specify vary depending on the command type and are shown below. However,
the total number of channels that you can specify varies depending on the model.
R-Math: C01 to C24, Write: 1 to 12, W-Math: 101 to 124

• Connected to (server number)

Select the server number from 1 to 16.

• Register

Set the register number of the server.
You can specify an input register in the range of 30001 to 39999 or 300001 to 365536.

You can specify a hold register in the range of 40001 to 49999 or 400001 to 465536.

The register numbers you can specify vary depending on the command type. See

section 6.3 in the FX1000 Communication Interface User’s Manual, IM 04L21B01-
17EN.

• Type

Select INT16, UINT16, INT32_B, INT32_L, UINT32_B, UINT32_L, FLOAT_B, or

FLOAT_L.

The types that you can specify vary depending on the type of command. See section 6.3

in the FX1000 Communication Interface User’s Manual, IM 04L21B01-17EN.