Input type (mode and range/type), Input type (mode and range/type) -6 – Yokogawa Value Series FX1000 User Manual
Page 19
2-6
IM 04L21B01-64EN
Buttons for Configuring the Selected Channels at Once
Copies the settings of the first channel
Set all channels at once.
Turns all channels ON or OFF
Initializes all channels
Set all values to their minimum settable values.
Set all values to their maximum settable values.
Input Type (Mode and Range/Type)
Delta, scaling, and square root are supported for the various modes as shown below.
Mode
Delta, Scaling, and Square Root
OFF
DELTA
SCALE
SQRT
LogType1
LogType2
SKIP
No
No
No
No
No
No
VOLT
Yes
Yes
Yes
Yes
Yes
Yes
TC
Yes
Yes
Yes
No
No
No
RTD
Yes
Yes
Yes
No
No
No
DI
Yes
Yes
Yes
No
No
No
1-5 V
No
No
Yes
No
No
No
The values in the Range/Type list box vary depending on the above settings.
• Span L, Span U
Input range. You cannot enter values that are outside of the range.
Note
• You cannot set the same value to [Span L] and [Span U].
• When [Delta/Scale/Sqrt/LOG Scale] is set to [Sqrt], [LogType1], or [LogType2], or [Mode] is
set to [1-5V], you can only set [Span L] to a value that is less than [Span U].
2.3 Setting the Measurement Channels