3 supplying power to the converter, Display functions -5 – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual
Page 94
IM 11M12A01-03E
7-2
7.3
Supplying Power to the Converter
CAUTION
To avoid temperature changes around the sensor, it is recommended that rather than
tuning it on and off power be continuously supplied to the High-temperature Humidity
Analyzer if it is used in an application where it is used periodically.
It is also recommended to flow a span gas (instrument air) beforehand.
Supply power to the converter. A display as in Figure 7.1, which indicates the detector’s
sensor temperature, then appears. As the heat in the sensor increases, the temperature
gradually rises to 750
Њ C. This takes about 20 minutes after the power is turned on,
depending somewhat on the ambient temperature and the measured gas temperature.
After the sensor temperature has stabilized at 750
Њ C, the converter is in measurement
mode. The display panel then displays the oxygen concentration as in Figure 7.2. This is
called the basic panel display.
Warmup
Tag:
319
Њ
C
4.00mA -Output1
4.00mA -Output2
Hold
Tag:
0.0
%H
2
O
4.00mA -Output1
4.00mA -Output2
F7.02E.EPS
Figure 7.1 Display of Sensor Figure 7.2 Measurement Mode Display
Temperature During Warmup