6 other settings, 1 setting the date-and-time, Which maximum and minimum values are monitored -19 – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual
Page 130: Setting purging -22
IM 11M12A01-03E
8-18
8.6
Other Settings
8.6.1
Setting the Date-and-Time
The following describe how to set the date-and-time. Automatic calibration or blowback
works following this setting.
Proceed as follows:
(1) Press the Setup key in the Basic panel display to display the Execution/Setup
display.
(2) Select Setup in the Execution/Setup display to display the “Commissioning” (Setup)
display.
(3) Select Others in the “Commissioning” (Setup) display. The Others display then
appears, as shown in Figure 8.15.
(4) Select Clock. The Clock display then appears, as shown in Figure 8.16.
(5) Select Set date to display the numeric-data entry display. To set the date June 21,
2000, enter 000621 and press the [Enter] key. The display then returns to the one
shown in Figure 8.16.
(5) Select Set time. Enter the time on a 24-hour basis. To enter 2:30 p.m., type in 1430
in the numeric-data entry display. Press the [Enter] key, and the time starts at 00
second.
Clock
᭜
Averaging
᭜
Fuel setup
᭜
Purging
᭜
Unit setup
᭜
Passwords
᭜
Defaults
Enter
Others
F8.13E.EPS
Enter
Set date: 2 1 / 0 6 / 0 0
᭜
Set time: 1 4 : 3 9
2 1 Jun. 2 0 0 0
1 4 : 3 0 : 0 0
Clock
F8.16E.EPS
Figure 8.15 Other Settings Figure 8.16 Clock Display