IM 11M12A01-03E
3-3
3. Installation
3.1.3
Installation of the Detector
CAUTION
• The cell (sensor) at the tip of the detector is made of ceramic (zirconia). Do not drop
the detector, as impact will damage it.
• A gasket should be used between the flanges to prevent gas leakage. The gasket
material should be heatproof and corrosion-proof, suited to the characteristics of the
measured gas
The following should be taken into consideration when mounting the general-use
detector:
(2) Where the detector is mounted horizontally, the calibration gas inlet and the refer-
ence gas inlet should face downward.
3.1.4
Installation of ZH21B Dust Protector
(1) Put the gasket that is provided by the user between the flanges and mount the dust
protector in the probe insertion hole.
(2) Make sure that the cell assembly mounting screws (four) at the probe tip are not
loose.
(3) Mount the detector so that the calibration gas inlet and the reference gas inlet face
downward
F3.2E.EPS
Unit
Ϻ
mm
Reference gas inlet
Calibration gas inlet
Figure 3.2 Installation of the dust filter