8 setting display item, 2 checking calibration contact output -14 – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual
Page 102
IM 11M12A01-03E
7-10
7.8
Setting Display Item
This section briefly describes the display item settings shown in Figure 7.11, “Basic
Panel Display.” If the humidity analyzer /HS option was specified at the time of pur-
chase, the primary value has set “Humidity.” If a mixing ratio is to be measured, change
the current primary value following the steps below.
Additionally, if the “humidity” has been selected in the Detector Type Setting in Section
7.6, earlier in this manual, the primary value is set to the humidity and the secondary
and tertiary values are current outputs 1 and 2, respectively when data initialization is
performed.
F7.11E.EPS
Tag:
0.0
%H
2
O
4.00mA -Output1
4.00mA -Output2
Tag name
Primary value
Secondary value
Tertiary value
Figure 7.11 Basic Panel Display
(1) Press the Setup key in the basic panel display to display the Execution/Setup display.
Then select Maintenance in the Execution/Setup display.
(2) Select the Display setup from the Maintenance panel display (Figure 7.12). The
Display setup display (Figure 7.13) then appears.
(3) In the above Display setup display, select the Display item. The Display item display
(Figure 7.14) then appears. From this display, select the Primary value and press the
[Enter] key to display the Display item selection display (Figure 7.15).
(4) Select the Secondary and Tertiary values in the same manner as in the steps above.
(5) Consult Table 7.2, “Display Items,” enabling the selection of display items in
individual display areas.
Enter
Display setup
᭜
Calibration setup
᭜
Blow back setup
᭜
mA-output loop check
᭜
Contact check
Maintenance
F7.12.EPS
Enter
Display item
᭜
Trend graph
᭜
Auto return time: 0min
᭜
Language: English
Display setup
F7.13.EPS
Figure 7.12 Maintenance Panel Display Figure 7.13 Display Setup