Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual
Page 162
IM 11M12A01-03E
10-14
Table 10.5 Initialization Items and Default Values
T10.5E-1.EPS
Item
Data to be initialized
Default setting
Type of equipment
Not initialized
Detector
ZR22
Measurement gas
Wet gas
Display item
1st display item
Humidity
2nd display item
Current output 1
3rd display item
Current output 2
Tag name
Deleted
Displayed data
Trend graph
Parameter
Humidity
Sampling interval
30 seconds
Upper limit (graph)
25% O
2
Lower limit (graph)
0% O
2
Automatic return time
0 min.
Language
Not initialized
Mode
Manual
Calibration procedure Span - zero
Zero-gas concentration 1.00% O
2
Span-gas concentration 21.00% O
2
Calibration data
Calibration setting Output hold time
10 min., 00 sec.
Calibration time
10 min., 00 sec.
Interval
30 days, 00 hr.
Start date
10/01/00
Start time
00:00
Mode
No function (invalid)
Blow back Blow back setting (Output) hold time
10 min., 00 sec.
Blow back time
10 min., 00 sec.
Interval
30 days, 00 hr.
Start date
01/01/00
Start time
00:00
Parameter
Humidity
mA-output 1
Min. oxygen concentration 0% O
2
mA-output 2
Max. oxygen concentration 25% O
2
Output damping
0%
Output mode
Linear
Warm-up
4 mA
Set value
4 mA
Maintenance
Previous value held
Output hold setting Set value
4 mA
Calibration, blowback
Previous value held
Set value
4 mA
Error
Preset value held
Set value
3.4 mA
Equipment
selection
Current output
data
to be continued next page