3 operational data initialization, Handling of the zo21s standard gas unit -18, Warning – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual
Page 161
IM 11M12A01-03E
10-13
10. Other Functions
10.3 Operational Data Initialization
Individual set data initialization enables you to return to the default values set at the time
of delivery. There are two types of initialization: an all set-data initialization and a
function-by-function initialization. Table 10.5 lists the initialization items and default
values.
To initialize the set data, follow these steps:
(1) Press the Setup key in the Basic panel display to display the Execution/Setup display.
Then choose Setup.
(2) Select Others from the “Commissioning” (Setup) display.
(3) Choose “Defaults.” The “Defaults” display shown in Figure 10.11 appears.
(4) Select your desired item to initialize in order to display the “Defaults” Display
shown in Figure 10.12.
(5) Choose Defaults start. Initialization then starts.
Display
᭜
Calibration
᭜
Blow back
᭜
mA-outputs
᭜
Alarms
᭜
Contacts
᭜
Others
᭜
All data
Enter
Defaults
F10.11E.EPS
Enter
Defaults start
᭜
Cancel
Defaults
Display
Figure 10.11 Defaults Display Figure 10.12 Initialization Start Display
WARNING
• Do NOT turn off the power during initialization. Otherwise, initialization will not be
performed properly.