3 output hold setting, 4 default values, Alarm setting -9 – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual
Page 120: Alarm values -9, Ma-outputs presets
IM 11M12A01-03E
8-8
8.2.3
Output Hold Setting
To set the output hold, follow these steps:
(1) Press the Setup key in the basic panel display to display the Execution/Setup display.
Then select Setup in the Execution/Setup display. Next, select the mA-output setup
and then the mA-output preset display as shown in Figure 8.2.
Warm up: 4mA
᭜
Preset value: 4 . 0 mA
᭜
Maintenance: Hold
᭜
Preset value: 4 . 0 mA
᭜
Cal.blowback: Hold
᭜
Preset value: 4 . 0 mA
᭜
Error: Preset
᭜
Preset value: 3 . 4 mA
Enter
mA-outputs presets
F8.2E.EPS
Warm up: 4mA
᭜
Preset value: 4 . 0 mA
᭜
Maintenance: Hold
᭜
Preset value: 4.0mA
᭜
Cal.blowback: Hold
᭜
Preset value: 4 . 0 mA
᭜
Error: Preset
᭜
Preset value: 3 . 4 mA
Enter
mA-outputs presets
Non-Hold
Hold
Preset
F8.3E.EPS
Figure 8.2 mA-output Preset Display Figure 8.3
(2) From the this display
͑Fig.8.2͒, select the desired display. Figure 8.3 shows an
example of selecting Maintenance. Select the desired output status.
(3) If a preset value is selected, set the corresponding output current. If you select a
preset value just below Maintenance, on the screen, the numeric-data entry display
will appear. Enter the current value you want. To set 10 mA, type in 100 and press
the [Enter] key to complete the setting.
The setting range is from 2.4 to 21.6 mA
8.2.4
Default Values
When the analyzer is delivered, or if data are initialized, output hold is the default as
shown in Table 8.3.
Table 8.3 Output Hold Default Values
Status
Output hold (min. and max. values)
During warm-up
4 mA
Under maintenance
Holds output at value just before maintenance started.
Under calibration or blow back
Holds output at value just before starting calibration
or "blow back."
On Error occurrence
Holds output at a preset value.
T8.3E.EPS
Preset value
4 mA
4 mA
4 mA
3.4 mA