4 entering tag name, 5 language selection, Blowback -9 – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual
Page 156: 1 blowback setup -9
IM 11M12A01-03E
10-8
10.1.4
Entering Tag Name
You can attach a desired tag name to the equipment. To attach it, follow these steps:
(1) Select the Setup key from the basic panel display to display the Execution/Setup
display. Then select Maintenance from the Execution/Setup display.
(2) Select the Display setup from the Maintenance display.
(3) Select the Display item from the Display setup display. The display shown in Figure
10.6 then appears.
(4) Select the Tag name from the Display item. The text-data entry display then appears.
(5) Enter up to 12 alphanumeric characters including codes for the desired tag name.
Enter
᭜
Primary value:
Humidity
᭜
Secondary value:
mA-ouput1
᭜
Tertiary value:
mA-output2
Tag name:
Display item
F10.6E.EPS
Figure 10.6 “Display Item” Display
10.1.5
Language Selection
You can select either English, German, or French as the language for the display. If you
selected an English display (basic code: -E, for the display) at the time of purchase,
“English” has already been selected for the language.
To select the language you want, follow these steps:
(1) Select the Setup key from the basic panel display to display the Execution/Setup
display. Then select Maintenance from the Execution/Setup display.
(2) Select the Display setup from the Maintenance display.
(3) Select Language from the Display setup display. The language selection display as
shown in Figure 10.7 then appears.
Enter
᭜
Display item
᭜
Trend graph
᭜
Auto return time: 0 min
Language:
Display setup
F10.7E.EPS
English
Deutsch
Francias
Figure 10.7 Display Setup Display