2setting periods which average, Setting passwords -23 – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual
Page 131
IM 11M12A01-03E
8-19
8. Detailed Data Setting
8.6.2
Setting Periods over which Average Values Are Calculated and Periods over
which Maximum and Minimum Values Are Monitored
The equipment enables the display of oxygen concentration average values and maxi-
mum and minimum values under measurement (see Section 10.1.1, later in this manual).
The following section describes how to set the periods over which oxygen concentration
average values are calculated and maximum and minimum values are monitored.
8.6.2.1Procedure
(1) Press the Setup key in the Basic panel display to display the Execution/Setup
display.
(2) Select Setup in the Execution/Setup display to display the Commissioning” (Setup)
display.
(3) Select Others in that display and then select Averaging in the Others display. The
averaging display shown in Figure 8.15 then appears.
(4) Choose “Set period over which average is calculated” and enter the desired numeric
value from the numeric-data entry display. To enter three hours, type in 003. The
period over which average values can be calculated ranges from 1 to 255 hours.
(5) Choose “Set period over which maximum and minimum is stored” and enter the
desired numeric value from the numeric-data entry display. To enter 48 hours, type
in 048. The allowable input ranges from 1 to 255 hours.
8.6.2.2Default Values
When the analyzer is delivered, or if data are initialized, the average-value calculation
periods and maximum- and minimum-value monitoring periods are by default one hour
and 24 hours respectively.
Set period over which
average is calculated:
1 h
᭜
Set period over which
maximum and minimum is
stored: 2 4 h
Enter
Averaging
F8.15E.EPS
Figure 8.17 Setting Average-Value Calculation Periods and Maximum- and
Minimum-Value Monitoring Periods