Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual
Page 21
IM 11M12A01-03E
2-2
Repeatability: (Note1)
Ϯ1 vol% H
2
O (sample gas pressure 2 kPa or less)
Linearity:
(Excluding standard gas tolerance)
(Note1)
(Use oxygen of known concentration (in the measuring range) as the
zero and span calibration gas.)
Ϯ2 vol% H
2
O; (Sample gas pressure: within
Ϯ0.49 kPa)
Ϯ3 vol% H
2
O; (Sample gas pressure: 2 kPa or less)
Drift:
(Excluding the first two weeks in use)
(Note1)
both zero and span
Ϯ3 vol% H
2
O/month
Response Time: Response of 90 % within 5 seconds. (Measured after gas is introduced
from calibration-gas inlet and analog output starts changing.)
(Note1) These tolerances do not apply to the pressure compensated
version, or where natural convection is used for the reference air.