<5. INSTALLATION AND WIRING> 5-21

IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00

5.11.3 Purging Analyzer and Universal Power Supply and/or URD for Hazardous
Areas (with On-Line Validation)

• NEC/CSA Class 1, Division 2, Groups A-D

• ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!

The block diagram below shows the sections of the analyzer that require nitrogen purging when using in
conjunction with either or the Universal Power Supply and Universal Remote Display. AZ-Type purge control
system is fitted the Launch Unit and it includes a local indicator (Bright Green, rugged light) and pressure switch
alarm contacts (open on loss of purge pressure).

5.11.4 Purging Analyzer and Universal Power Supply and/or URD (not using On-Line Validation)


• NEC/CSA Class 1, Division 2, Groups A-D

• ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!

The block diagram below shows the sections of the analyzer that require nitrogen purging. A Z-Type purge
control system is fitted the Launch Unit and it includes a local indicator (Bright Green, rugged light) and
pressure switch alarm contacts (open on loss of purge pressure). The purging should be carried in sequence
typically as shown below. All purge gas connections are ¼” od Tube fittings.

TDLS200 TDL Analyzer Instruction Manual V2.1

Page 49 of 131

- 49 -

3.9.2 Purging Analyzer for Hazardous Areas (without On-Line Validation)

Z-Purged designed in accordance with NEC/CSA Class 1, Division 2, Groups A-D

Certified ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!

The block diagram below shows the sections of the analyzer that require nitrogen purging.

A Z-Type purge control system is fitted the Launch Unit and it includes a local indicator

(Bright Green, rugged light) and pressure switch alarm contacts (open on loss of purge

pressure).

Figure 28 - Purge Flow Diagram when not using on line validation

3.9.3 Purging Analyzer and Universal Power Supply and/or URD for Hazardous

Areas (with On-Line Validation)

Z-Purged designed in accordance with NEC/CSA Class 1, Division 2, Groups A-D

Certified ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!

The block diagram below shows the sections of the analyzer that require nitrogen purging when

using in conjunction with either or the Universal Power Supply and Universal Remote Display. A Z-

Type purge control system is fitted the Launch Unit and it includes a local indicator (Bright Green,

rugged light) and pressure switch alarm contacts (open on loss of purge pressure).

Figure 29 - Purge Flow Diagram when using on line validation

Main

Electronic

Housing &

Purge

System

Laser

Module

Check Gas

Flow Cell

Proce

ss

Inte

rfa

ce

Proce

ss

Inte
rfa

ce

Detect

Module

Detect

Electronic

Housing

Nitrogen or I/A Purge

Gas

Nitrogen or I/A Purge

Gas

Power

Supply

or URD

Main

Electronic

Housing &

Purge

System

Laser

Module

Check Gas

Flow Cell

Pr

oce

ss

Int

erf

ac

e

Pr

oce
ss

Int

erf

ac

e

Detect

Module

Detect

Electronic

Housing

Nitrogen or I/A Purge Gas

Figure 49 - Purge Flow Diagram when using on line validation

Nitrogen or I/A Purge
Gas

Check Gas
Flow Cell

Detect
Electronic
Housing

TDLS200 TDL Analyzer Instruction Manual V2.1

Page 50 of 131

- 50 -

3.9.4 Purging Analyzer and Universal Power Supply and/or URD (not using On-Line

Validation)

Z-Purged designed in accordance with NEC/CSA Class 1, Division 2, Groups A-D

Certified ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!

The block diagram below shows the sections of the analyzer that require nitrogen purging. A Z-

Type purge control system is fitted the Launch Unit and it includes a local indicator (Bright Green,

rugged light) and pressure switch alarm contacts (open on loss of purge pressure). The purging

should be carried in sequence typically as shown below. All purge gas connections are ¼” od Tube

fittings.

Figure 30 - Purging Analyzer for Hazardous Areas -NO- On-Line Validation

Main

Electronic

Housing &

Purge

System

Laser

Module

Check Gas

Flow Cell

Process Interface

Process

Interface

Detect

Module

Detect

Electronic

Housing

Nitrogen or I/A Purge Gas

Power

Supply

or URD

Figure 50 - Purging Analyzer for Hazardous Areas -NO- On-Line Validation

Main
Electronic
Housing
& Pureg
System

Detect
Electronic
Housing

Power
Supply
or URD

Check Gas
Flow Cell