Validation and calibration – Yokogawa In-Situ Gas Analyzer TDLS200 User Manual
Page 107
8-6
8-7
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<8. VALIDATION AND CALIBRATION>
After entering cell pressure, cell temperature and cell length (28.6” or 72.6cm) the following option
will appear on the basis of two gas measurement option. Please toggle to either of the two gases
configured within the analyzer or as in the example below, there is no second measured gas
configured.
Note, after entering the desired span gas value, the following display will prompt you to
press 9 to actually perform the calibration or escape to cancel. Note, the concentration that is being
displayed and up-dated in the lower box is the currently measured value before the calibration is
performed. After calibration, please check the results and consult with Yokogawa if any questions or
concerns.