Installation – Yokogawa digitalYEWFLO (DY) User Manual
Page 53

INSTALLATION
IM 1F2B4-01-YIA
Page 57
•Max. Entity Parameters
Vmax = 28 Vdc
Imax = 93 mA
Pmax = 0.66W
Analog Output
Pulse Output
Ci
4 nF
62 nF
Li
0 mH
0 mH
•Installation Requirements
Vmax > Voc or Vt
Imax > Isc or It
Ca > Ci + Ccable
La < Li + Lcable
Notes:
1.
YF011 signal cable max length 20 meters.
2.
Maximum non-hazardous location voltage must not exceed 250v.
3.
Do not alter drawing without authorization from FMRC.
4.
Installation should be in accordance with ANSI/ISA RP12.6. "Installation of Intrinsically Safe
Systems for Hazardous (Classified) Locations" and the National Electrical Code (ANSI/NFPA
70).
5.
Approved dust-tight seals are required for Class II and III installations.
See also other documents in the category Yokogawa Sensors:
- EJA115 (85 pages)
- EJA120A (47 pages)
- EJA120A (79 pages)
- EJA130A (2 pages)
- EJA130A (4 pages)
- EJA120A (31 pages)
- EJA130A (47 pages)
- EJA120A (40 pages)
- EJA438 (5 pages)
- EJA120A (6 pages)
- EJA210A (70 pages)
- EJA130A (4 pages)
- EJA430A (78 pages)
- EJA210E (52 pages)
- EJA210E (89 pages)
- EJA210E (170 pages)
- EJX120A (4 pages)
- EJA210E (9 pages)
- EJX115A (55 pages)
- EJA210E (41 pages)
- EJA210E (96 pages)
- EJX910A (83 pages)
- EJX910A (9 pages)
- EJX910A (103 pages)
- FlowNavigator Software (163 pages)
- EJX910A (55 pages)
- EJX910A (175 pages)
- EJA530A (67 pages)
- EJA120A (83 pages)
- EJX530A (52 pages)
- EJA110E (4 pages)
- EJA110E (85 pages)
- EJX120A (85 pages)
- EJA118 (76 pages)
- EJX118A (64 pages)
- EJA438 (72 pages)
- EJA430E (85 pages)
- EJX430A (76 pages)
- EJA430E (7 pages)
- EJX430A (6 pages)
- EJX430A (40 pages)
- EJA430E (41 pages)
- EJA430E (96 pages)
- EJX438A (10 pages)
- ADMAG AXR (194 pages)