Yokogawa ADMAG AXF User Manual
Page 76

IM 01E20D01-01E
6-14
6. OUTLINE
General-Purpose Use, Remote Flowtube (electrode
structure code 2: replaceable electrode)
Pressure
MPa (psi)
2 (290)
1 (145)
– 0.1 (–14.5)
0
(32)
40
(104)
100
(212)
160
(320)
–10
(14)
130
(266)
Temperature
°C (°F)
25 to 200 mm (1.0 to 8.0 in.) (flange type, wafer type)
250, 300 mm (10, 12 in.) (flange type)
350, 400 mm (14 to 16 in.) (flange type)
250, 300 mm (10, 12 in.) (wafer type)
F18-3.EPS
General-Purpose Use, Integral Flowmeter (electrode
structure code 2: replaceable electrode)
2 (290)
1 (145)
– 0.1 (–14.5)
40
(104)
100
(212)
130
(266)
250, 300 mm (10, 12 in.) (flange type)
350, 400 mm (14 to 16 in.) (flange type)
250, 300 mm (10, 12 in.) (wafer type)
25 to 200 mm (1.0 to 8.0 in.) (flange type, wafer type)
Temperature
°C (°F)
Pressure
MPa (psi)
0
(32)
–10
(14)
F18-4.EPS
Note: For replaceable electrodes for fluid temperatures of
-10
°C (14°F)or less, please contact Yokogawa office.
Sanitary Type (electrode structure code 1: Non-
replaceable electrode)
F18-5.EPS
Temperature
°C (°F)
Pressure
MPa (psi)
1(145)
–0.1(–14.5)
15 to 125 mm (0.5 to 5.0 in.) (integral flowmeter)
160
(320)
130
(266)
15 to 125 mm (0.5 to 5.0 in.) (remote flowtube)
0
(32)
–10
(14)
Note: In case of 120 to 160
°C (248 to 320°F) of fluid
temperature, please select optional code GH.
Ceramics Lining
General-Purpose Use and Explosion proof Type,
Remote Flowtube (electrode structure code 1: Non-
replaceable electrode)
Pressure
MPa (psi)
Temperature
°C (°F)
2.5 to 50 mm (0.1 to 2.0 in.)
80 to 200 mm (3.0 to 8.0 in.)
4 (580)
2 (290)
–0.1 (–14.5)
–10
(14)
0
(32)
100
(212)
180
(356)
F19-1.EPS
General-Purpose Use and Explosion proof Type,
Integral flowmeter (electrode structure code 1: Non-
replaceable electrode)
Pressure
MPa (psi)
4 (580)
3.25 (471)
2 (290)
–0.1(–14.5)
2.5 to 50 mm (0.1 to 2.0 in.)
80 to 200 mm (3.0 to 8.0 in.)
–10
(14)
0
(32)
100
(212)
130
(266)
180
(356)
Temperature
°C (°F)
F19-2.EPS
*1:
For fluid temperature of the explosion proof type, refer
to descriptions of “HAZARDOUS AREA CLASSIFI-
CATION”.
Polyurethane Rubber Lining
General-Purpose Use and Submersible Type, Remote
Flowtube (electrode structure code 1: Non-replace-
able electrode)
F20-1.EPS
Pressure
MPa (psi)
4 (580)
–10
(14)
0
(32)
40
(104)
Temperature
°C (°F)
2 (290)
1 (145)
–0.1(–14.5)
0.75(109)
25 to 50 mm (1.0 to 2.0 in.) (flange type, wafer type)
65 to 200 mm (2.5 to 8.0 in.) (flange type, wafer type)
250, 300 mm (10, 12 in.) (flange type)
250, 300 mm (10, 12 in.) (wafer type)
350 to 1000 mm (14 to 40 in.) (flange type)
1100 to 2600 mm (44 to 104 in.) (flange type)
General-Purpose Use, Integral Flowmeter (electrode
structure code 2: replaceable electrode)
F20-2.EPS
Pressure
MPa (psi)
2 (290)
1 (145)
–0.1(–14.5)
25 to 200 mm (1.0 to 8.0 in.) (flange type, wafer type)
250, 300 mm (10, 12 in.) (flange type)
350, 400 mm (14 , 16 in.) (flange type)
250, 300 mm (10 , 12 in.) (wafer type)
–10
(14)
0
(32)
40
(104)
Temperature
°C (°F)