Yokogawa ADMAG AXF User Manual
Page 55

IM 01E20D01-01E
5-6
5. MAINTENANCE
CAUTION
• The lining of sanitary type uses fluorocarbon
PFA. For the property of fluorocarbon PFA, it is
possible that the adapter mounting screws may
loosen as time passes, so retighten them
regularly.
• Be sure to retighten the adapter mounting
screws according to the prescribed torque
values in Table 5.3.1 or Table 5.3.2. Retighten
them diagonally with the same torque values,
step by step up to the prescribed torque value.
• In case of leakage from adapter connection
point, retighten the adapter mounting screws. If
leakage doesn’t stop even if they are retight-
ened, replace the gasket between mini-flange
of flowmeter and adapter.
• It is recommended to replace the gasket
periodically,
• It is necessary to decide the period between
changes with consideration for the frequency of
cleaning cycles, the cleaning temperature and
the fluid temperature.
• Use the gasket specified by Yokogawa in Table
5.3.3
Table 5.3.3 Parts number of gasket
T0504.EPS
EPDM rubber
(standard)
Silicon rubber
(optional
code GH)
15 (0.5)
25 (1.0)
32 (1.25)
40 (1.5)
50 (2.0)
65 (2.5)
80 (3.0)
100 (4.0)
125 (5.0)
F9811QA
B1002EG
B1003EG
B1004EG
B1005EG
B1006EG
B1007EG
B1008EG
F9811QC
F9811QB
B1010EG
B1011EG
B1012EG
B1013EG
B1014EG
B1015EG
B1016EG
F9811QD
Size mm
(inch)
Gasket
type