Page 52

IM 01E20D01-01E
5-3
5. MAINTENANCE
(6) Move aside the electrode lead wire and avoiding
the screw, insert end A of the tool to the electrode
and loosen the electrode holder using the torque
driver.
F0506.EPS
Torque driver
End A
(7) Screw end B of the tool into the screw block of the
electrode, and pull the tool straight out to remove
the electrode holder.
F0507.EPS
End B
5.2.2
Cleaning Replaceable Elec-
trodes
(1) Clean the electrode surface (wetted part) with
alcohol or other cleaning agents.
O-ring
Electrode surface
Electrode holde
F0508.EPS
NOTE
The screw threads are coated with an anti-
seizing compound. Do not let the compound
come in contact with the electrode.
(2) Clean the surface of the lining seal with an alcohol
moistened swab.
F0509.EPS
Lining seal surface
See also other documents in the category Yokogawa Sensors:
- EJA120A (79 pages)
- EJA130A (2 pages)
- EJA130A (4 pages)
- EJA120A (31 pages)
- EJA130A (47 pages)
- EJA120A (40 pages)
- EJA438 (5 pages)
- EJA120A (6 pages)
- EJA115 (85 pages)
- EJA120A (47 pages)
- EJA210A (70 pages)
- EJA130A (4 pages)
- EJA430A (78 pages)
- EJA210E (89 pages)
- EJA210E (170 pages)
- EJX120A (4 pages)
- EJA210E (9 pages)
- EJX115A (55 pages)
- EJA210E (41 pages)
- EJA210E (96 pages)
- EJA210E (52 pages)
- EJX910A (9 pages)
- EJX910A (103 pages)
- FlowNavigator Software (163 pages)
- EJX910A (55 pages)
- EJX910A (175 pages)
- EJX910A (83 pages)
- EJA530A (67 pages)
- EJA120A (83 pages)
- EJX530A (52 pages)
- EJA110E (4 pages)
- EJA110E (85 pages)
- EJX120A (85 pages)
- EJA118 (76 pages)
- EJX118A (64 pages)
- EJA438 (72 pages)
- EJA430E (85 pages)
- EJX430A (76 pages)
- EJA430E (7 pages)
- EJX430A (6 pages)
- EJX430A (40 pages)
- EJA430E (96 pages)
- EJA430E (41 pages)
- EJX438A (10 pages)
- ADMAG AXR (194 pages)