beautypg.com

2 adjusting zero point for static pressure, 3 adjusting zero point for external temperature, Adjusting zero point for static pressure -3 – Yokogawa EJX930A User Manual

Page 50: Adjusting zero point for external temperature -3

background image

<8. Operation>

8-3

IM 01C25R01-01E

The zero-adjustment screw is located inside the

cover.

Use a slotted screwdriver to turn the zero-

adjustment screw. Equalize the transmitter, then

turn the screw clockwise to increase the output or

counterclockwise to decrease the output. The zero

point adjustment can be made with a resolution

of 0.01% of the setting range. The degree of zero

adjustments varies with the screw turning speed;

turn the screw slowly to make a fine adjustment,

quickly to make a rough adjustment.

8.2.2 Adjusting Zero Point for Static

Pressure

Zero point adjustment for static pressure can

be adjusted with the HART Communicator or

FlowNavigator.

(1) When you can obtain the Low Range Value

from the actual measured value of 0% (0

MPa pressure);

Apply zero pressure in % to the transmitter. After

obtaining a stable pressure, adjust the zero point.

(2) When you cannot obtain the Low Range

Value from the actual measured value of

0%;

Adjust the transmitter output to the actual measured

value obtained by a highly accurate master meter

such as digital manometer.

Example:

Range: 0 to 16 MPa abs

Actual measured value: 2.000 MPa abs

Transmitter s output: 2.1 MPa abs

Lower SP trim deviation

(SP Manual Lower Pt): – 0.3 MPa abs

F0804.ai

EJX-MV:YOKOGAWA

SP trim mode:

1 Off

2 Auto, Lower Pt

3 Auto, Upper Pt

4 Manual Lower Pt

5 Manual Upper Pt

ABORT ENTER

(ENTER)

‘-0.4’

EJX-MV:YOKOGAWA

SP for trim 0.2100

manual, Lower Pt

-0.3000 MPa

-0.4

HELP

ABORT ENTER

SP for trim is 2.1000

Master meter shows 2.0 MPa abs

= 2.0-2.1

= -0.1

< Calculation of correction value>

= The deviation value for SP zero adjustment

+ error value for SP zero adjustment

= (-0.3)+(-0.1)

=-0.4

1. Device setup
2. Diag/Service
3. Calibration
4. SP sensor trim
1. SP trim

ENTER (F4)

×3

1

2

2

DEL

8.2.3 Adjusting Zero Point for External

Temperature

Zero point adjustment for external temperature

can be adjusted by the HART Communicator or

FlowNavigator.
As defined the reference resistor value table of

the thermometer resistor (RTD), obtain resistance

value corresponding to 0%, and use the obtained

resistance as the input value, then deliver it to the

transmitter by means of a variable resistor.
Measure the resulting output signal with the

voltmeter (digital multimeter), and check the output

value relative to the input value.
Zero point can be adjusted by the HART

Communicator's easy key operation.

This manual is related to the following products: