beautypg.com

2 wiring installation, 6 rtd cable connection, Wiring installation -4 – Yokogawa EJX930A User Manual

Page 41: Rtd cable connection -4 7.6.1, Connecting shielded cable with cable gland

background image

<7. Wiring>

7-4

IM 01C25R01-01E

7.5.2 Wiring Installation

(1) General-use Type and Intrinsically Safe

Type

With the cable wiring, use a metallic conduit or

waterproof glands.

• Apply a non-hardening sealant to the terminal

box connection port and to the threads on the

flexible metal conduit for waterproofing.

F0710.ai

Flexible metal conduit

Wiring metal

conduit

Tee

Drain plug

Apply a non-hardening

sealant to the threads

for waterproofing.

Figure 7.8

Typical Wiring Using Flexible Metal

Conduit

(2) Flameproof Type
Wire cables through a flameproof packing adapter,

or use a flameproof metal conduit.
■ Wiring cable through flameproof packing adapter.

• Apply a non-hardening sealant to the terminal

box connection port and to the threads on the

flameproof packing adapter for waterproofing.

Flameproof packing adapter

Flexible metal conduit

Wiring metal

conduit

Tee

Drain plug

Apply a non-hardening

sealant to the threads

for waterproofing.

F0711.ai

Figure 7.9

Typical Cable Wiring Using Flameproof

Packing Adapter

■ Flameproof metal conduit wiring

• A seal fitting must be installed near the terminal

box connection port for a sealed construction.

• Apply a non-hardening sealant to the threads of

the terminal box connection port, flexible metal

conduit and seal fitting for waterproofing.

F0712.ai

Non-hazardous area

Hazardous area

Flameproof

heavy-gauge

steel conduit

Tee

Drain plug

Seal fitting

Gas sealing device

Flameproof flexible

metal conduit

Apply a non-hardening

sealant to the threads

of these fittings for

waterproofing

After wiring, impregnate the fitting

with a compound to seal tubing.

Figure 7.10 Typical Wiring Using Flameproof Metal

Conduit

7.6 RTD Cable Connection

Connection of the RTD cable is always required

to measure external temperature. Follow the

procedures below to connect and disconnect a

cable when a cable gland or a conduit is used.

7.6.1 Connecting Shielded Cable with

Cable Gland

(External temperature input code:

-1, -2, -3, and -4)

• RTD connection components: EJX multivariable

transmitter, two cable glands, and RTD cable

Two cable glands are attached.

Y

L

P

P

U

S

ESL

UP

K

C

E

H

C

M

R

A

L

A

Transmitter

Cable gland

Electrical connection

RTD cable

with connector

Detail of connector

F0713.ai

This manual is related to the following products: