<10. General Specifications>
10-1
IM 01C25T04-01EN
10. General Specifications
10.1 Standard Specifications
For items other than those described below,
refer to each User’s Manual.
Applicable Model:
All DPharp EJX/EJA series.
Output:
Digital communication signal based on
PROFIBUS PA protocol.
Supply Voltage:
9 to 32 V DC for general use, flameproof type,
intrinsically safe (Ex ic), or nonincendive
9 to 24 V DC for intrinsically safe type Entity
model
9 to 17.5 V DC for intrinsically safe type FISCO
model
Communication Requirements:
Supply Voltage: 9 to 32 V DC
Current Draw: 15 mA (max)
Response Time (for Primary Value):
F, L capsule: 185 ms
M, H, A, B, C, D capsule: 150 ms
When amplifier damping is set to zero, and
including dead time.
Update Period:
Differential Pressure: 100 ms
Static Pressure: 100 ms
Capsule Temperature: 1 s
Amplifier Temperature: 1 s
Integral Indicator (LCD display):
5-digit Numerical Display, 6-digit Unit Display
and Bar graph. The indicator is configurable
to display one or up to four of the I/O signals
periodically.
Functional Specifications:
Functional specifications for PROFIBUS
communication conform to the PROFIBUS-PA
ver 3.02
Function Block:
Three AI blocks, One Totalizer block.