2 sensor transducer block, Sensor transducer block -3 – Yokogawa EJX115A User Manual
Page 63
<9. Parameter Lists>
9-3
IM 01C25T04-01EN
Relative
Index
Index
Parameter
Write
Mode
Initial Value
Explanation
72
88
PRIVATE_9
—
—
Not used for the transmitter
73
89
PRIVATE_10
—
—
Not used for the transmitter
74
90
PRIVATE_11
—
—
Not used for the transmitter
9.2 SENSOR Transducer Block
Relative
Index
Index
Parameter
Write
Mode
Initial Value
Explanation
0
16
BLOCK_OBJECT
—
—
Information on this block such as Block Tag, DD Revision,
Executiion Time etc.
1
17
ST_REV
—
0
The revision level of the static data associated with the
function block. The revision value will be incremented each
time a static parameter value in the block is changed.
2
18
TAG_DESC
AUTO
Space
The user description of the intended application of the
block.
3
19
STRATEGY
AUTO
0
The strategy field can be used to identify grouping of blocks.
This data is not checked or processed by the block.
4
20
ALERT_KEY
AUTO
0
The identification number of the plant unit. This information
may be used in the host for sorting alarms, etc.
5
21
TARGET_MODE
AUTO
AUTO
Set the Target of block mode (MODE_BLK) to Auto or O/S
according to the write Mode of the parameter to be set or
changed.
6
22
MODE_BLK
—
AUTO
The mode parameter is a structured parameter composed
of the actual mode, the normal mode, and the permitted
mode.
7
23
ALARM_SUM
—
—
The current alarm status associated with the function block
8
24
SENSOR_VALUE
—
0
Raw sensor value. The uncalibrated measurement value
from the sensor.
9
25
SENSOR_HI_LIM
—
100
Sensor upper limit value.
10
26
SENSOR_LO_LIM
—
-100
Sensor lower limit value.
11
27
CAL_POINT_HI
O/S
100
The highest calibrated value.
12
28
CAL_POINT_LO
O/S
0
The lowest calibrated value.
13
29
CAL_MIN_SPAN
—
0.5
The minimum calibration span value allowed.
14
30
SENSOR_UNIT
O/S
1133: kPa
Engineering unit's index code for the calibration value.
SENSOR_UNIT must be a subset of the interchangeable
part of the Pressure unit. Please refer to ordering instruction
for the acceptable units.
15
31
TRIMMED_VALUE
—
—
Sensor value after the trim processing.
16
32
SENSOR_TYPE
—
126: Resonant
The type of sensor
17
33
SENSOR_SERIAL_
NUMBER
—
0
The serial number of sensor
18
34
PRIMARY_VALUE
—
—
Measured value and status available to the Function Block.
19
35
PRIMARY_VALUE_
UNIT
O/S
kPa
Engineering unit code for the primary value according to
PRIMARY_VALUE_TYPE.
20
36
PRIMARY_VALUE_
TYPE
O/S
0: Pressure
This parameter contains the application of the pressure
device.
0: Pressure
1: Flow
2: Level
3: Volume
128:differential pressure
129:gauge pressure
130:absolute pressure
253: Special
21
37
SENSOR_
DIAPHRAGM_
MATERIAL
O/S
Unknown
Index code for the material of the diaphragm, which comes
in contact with the process media.
22
38
SENSOR_FILL_
FLUID
O/S
Unknown
Index code for the fill fluid inside the sensor. The index code
is manufacturer specific.
24
40
SENSOR_O_RING_
MATERIAL
O/S
Unknown
Index code for the material of the o-ring between the
diaphragm and process connection.