background image

CATC

CATC Inspector User’s Manual

Version 2.31

20

a static bus event (Suspend, Resume, Reset)

an external trigger signal input

various combinations of the above

Selection of the “Event Trigger” recording mode will enable the “Setup...” button immediately to
its right; clicking on the button will bring up the following dialog:

This dialog box contains four panels and a checkbox; the following examines each of these in
turn...

4.4.1.1.4.1 Packet Events

This panel allows for the selection of a variety of packet types (and, in some cases, contents). The
detection of a packet matching the specified criteria will be interpreted as a trigger event.

PID Type: This is the default event selection (using the dialog box shown above); it permits the
operator to select a particular PID type to act as the Event Trigger. Two of the permitted options
(“IN, OUT, or SETUP” and “DATA0 or DATA1”) combine multiple PID types into one
selection. A checkbox (“Trigger on any PID type NOT selected”) inverts the meaning of the PID
selection: the chosen item will be the only one not interpreted as a trigger event.