beautypg.com

Display formats, Display setup, Wavejet 300a series – Teledyne LeCroy WaveJet 300A Series Getting Started Guide User Manual

Page 55

background image

WaveJet 300A Series

54

WJ-A-GS-E Rev A

Other Parameters

Integral -- Computes area of waveform between cursors relative to zero level.

Values greater than zero contribute positively to the area; values less than zero
negatively.

Skew -- Measures from the 50% crossing of the first edge of a channel to the

50% crossing of a second channel.

Skew@Level -- Same as Skew, but with user-defined level.

Statistics

For each parameter, you can display minimum and maximum values by pressing the
M

EASURE

button, then setting Min/Max to On.