Teledyne LeCroy LabMaster 9Zi-A User Manual
Page 189

Getting Started Manual
922157-00 Rev A
181
1. Touch Analysis → Pass/Fail Setup... from the menu bar.
2. Touch one of the Qx tabs and a setup dialog for the position is
shown.
3. Touch inside the Condition field in the main dialog and select
DualParamCompare.
4. Touch inside the Source1 and Source2 fields and select a source
from the pop-up menu.
5. Touch inside the ParamCompare mini-dialog field and select a
source from the pop-up menu.
6. Touch inside the Compare Values field and select All or Any from
the pop-up menu.
By selecting All, the test are true only if every waveform falls within
the set limit. Selecting Any makes the test true if just one waveform
falls within the limit.
7. Touch inside the Condition field on the ParamCompare part of the
dialog and select a math operator from the pop-up menu.
8. Touch inside the Limit field and enter a value, using the pop-up
numeric keypad. This value takes the dimension of the parameter
that you are testing. For example, if you are testing a time
parameter, the unit is seconds.
9. If you choose either WithinDeltaPct or WithinDeltaAbs from the
Condition menu respectively, touch inside the Delta field and enter
a value.