If you perform ChannelSync Cal on multiple gain settings, perform
it first on the 100 mV/div setting, then perform it on other gain
settings (with “Reference Correction” values entered as
appropriate for those other gain settings.)
For the highest possible accuracy across the full input range of the
oscilloscope, a ChannelSync Cal should be performed on at least
one gain setting in each of the following gain ranges:
13-20 GHz ProLink
"A" Inputs
30-45 GHz 2.92mm or 2.4mm
"B" Inputs
<11.8 mV/div
10 mV/div
11.9 to 17.0 mV/div
10.1 to 12.6 mV/div
17.1 to 20.5 mV/div
12.7 to 15.8 mV/div
20.6 to 30.0 mV/div
15.9 to 20.0 mV/div
30.1 to 39.6 mV/div
20.1 to 25.0 mV/div
39.6 to 50.0 mV/div
25.1 to 31.5 mV/div
50.1 to 90.0 mV/div
31.6 to 39.5 mV/div
90.1 to 100.0 mV/div
39.6 to 50.0 mV/div
100.1 to 118.0 mV/div
50.1 to 63.0 mV/div
118.1 to 170.0 mV/div
63.1 to 79.0 mV/div
170.1 to 205.0 mV/div
79.1 to 82.0 mV/div
205.1 to 300.0 mV/div
82.1 to 100.0 mV/div
300.1 to 395.0 mV/div
100.1 to 126.0 mV/div
395.1 to 500.0 mV/div
126.1 to 158.0 mV/div
500.1 to 900.0 mV/div
158.1 to 200.0 mV/div
900.1 to 1000 mV/div
200.1 to 250.0 mV/div
If you add cables or probes to your setup subsequent to running
the ChannelSync Cal, you must manually deskew these items in the
Channel dialog to account for their propagation delays. The values
you provide on the Channel dialog (in the deskew section) are
additional to values shown on the ChannelSync Calibration dialog.