Getting Started Manual
922157-00 Rev A
105
4. Touch the Enable Timeout checkbox.
5. Touch inside the Timeout data entry control and provide a timeout
value.
NOTE:
Use the sequence mode timeout to automatically interrupt the
sequence acquisition if the timeout value is exceeded without a valid
trigger. The timeout period accounts for instances when a Num Segments
miscount occurs for some reason and the oscilloscope waits indefinitely for
an unforthcoming segment. During that time, no oscilloscope functions are
accessible. By means of a timeout value, however, the acquisition will be
completed, the waveform displayed, and control of the oscilloscope
returned to the user after the timeout has elapsed.
5. Under Display Settings, touch inside the Display mode control, and
select a sequence mode display from the pop-up menu.
6. Touch the S
INGLE
trigger front panel button.
NOTE:
Once a single acquisition has started, you can interrupt the
acquisition at any time by pressing the
S
INGLE
front panel button a second
time or by pressing the
S
TOP
front panel button. In this case, the segments
already acquired will be retained in memory.