9060-Zirconia

Analyser


Configuration and Setup Addresses

Holding

Reg. Function

Description

716 Probe #1 offset

10 = 1.0mV

717 Probe #2 offset

10 = 1.0mV

Purge control related variables

754 Purge enable

0= off, 1= on

Calibration checking gas related variables

759 Gas calibration check

0= off, 1= 1 gas, 2= 2 gasses

204
8

Probe #1 EMF

100,000 = 100.000 mV

205
0

Probe #2 EMF

100,000 = 100.000 mV

205
2

Probe #1 OXYGEN

100,000,000 = 100.0%

205
4

Probe #2 OXYGEN

100,000,000 = 100.0%

205
6

Probe #1, Impedance

1,000 = 1 k

Ω

205
8

Probe #2, Impedance

1,000 = 1 k

Ω

206
0

Probe #1 TC mV

100,000 = 100.000 mV

206
2

Probe #2 TC mV

100,000 = 100.000 mV

206
4

Probe #1 temperature

700 = 700 degC

206
6

Probe #2 temperature

700 = 700 degC

206
8

ALRM-ARRAY

Array of current alarm status. See below

208
4

ALRM-TIMES

Array of timestamp of alarms


Alarm array order -


1. Heater 1 fail
2. Sensor1 fail (Impedance too high)
3. Probe 1 filter blocked
4. Probe 1 thermocouple open circuit
5. Reference air Pump fail
6. Battery backed RAM fail
7. Mains frequency measurement fail
8. ADC warning (outside normal specifications, but still accurate)
9. DAC warning (outside normal specifications, but still accurate)