9060-zirconia analyser – Teledyne 9060 - Zirconium oxide flue gas oxygen analyzer User Manual

Page 30

background image

9060-Zirconia

Analyser

Sample Piping -

Connect the gas sample piping to the “sample in” port. If the process,

boiler, kiln or furnace has a positive pressure, no suction will be required. If the sample is
under a negative pressure, connect a pump to the “inlet” port as shown below. The flow
rate should be within the range of 1 to 5 litres/minute (120 to 600 scfm).












3.3

INSTALLING THE AUXILIARY THERMOCOUPLE


Weld a 1/2 inch BSP mounting socket to the flue within about 300 mm (12”), and

upstream of the oxygen probe. The thermocouple should be of similar length to the
oxygen probe to prevent flue temperature distribution errors.


3.4

SHIELD CONNECTIONS

All external wiring to the 9060 Analyser should be shielded. Do not connect shields

at the field end. Simply clip off and insulate. An extra terminal strip may be required to
connect all shields together. This should be supplied by the installer.


Optional vent to
atmospherre

Optional return
to flue

Upward sloping
sample line 3/8”
stainless steel

Flue
Gas

Flowmeter

Sampling
Probe

1234
Sensor

Vent or return
to process

Dry
Process
Gas

1/4" Stainless Steel Tube

1234
Sensor