Teledyne 9060 - Zirconium oxide flue gas oxygen analyzer User Manual
Page 59
9060-Zirconia
Analyser
59
59
NOTE:
If output channel is selected for oxygen, and the temperature of sensor 1 or
sensor 2 is below 650
°C (the display will be flashing “NOT READY”) the 4-20mA output for
the associated channel will be set to zero mA.
See set-up 91 for further details.
27. Transmitter Span Channel 1
Applicable only to linear outputs. Select transmitter span for output Channel 1. For
combustion applications, typical linear spans are 0 to 10 % or 0 to 15 % oxygen. Very low
oxygen range is adjustable from 0 to 0.001, to 2.000%.
Default setting is 10.0%.
28. Transmitter Output Channel 2
Select transmitter output for output Channel 2.
Options:
Sensor EMF 1
*
1. Logarithmic oxygen, 0.1 to 20 %
2. Oxygen deficiency, sensor 1 (or Sensor 2 if 2 sensors are selected in set-up 1)
3. % carbon dioxide dry
4. Auxiliary (Flue) temperature
5. Linear oxygen % sensor 1 ( or Sensor 2 if 2 sensors are selected in set-up 1)
6. 1 x 10 +2 to 10-30 % oxygen sensor 1 (or Sensor 2 if 2 sensors are selected in set-up
1), for reducing conditions.
7. Combustibles %
8. Linear Oxygen Average (Sensor 1 & 2) (When 2 sensors are selected in set-up 1)
29. Transmitter Zero Channel 2
The output zero and span of Channel 2 is set in set-up steps 29 and 30. Range limits are
shown below.