Teledyne 9060 - Zirconium oxide flue gas oxygen analyzer User Manual

Page 53

background image

9060-Zirconia

Analyser

53

53

82 Data to Print

83 Print Log Period

84 Printer Baud Rate


85 Reference air pump Internal/External/Inst air

86 Reference air RH if ‘Instrument Air’ selected in Set-up 85.

87

Damping

factor

88

MODBUS™

Address

The “Extended Set-up Menus” steps 89 to 97 will be skipped unless the extended menus
are enabled. See Set-up 5.2 for more details.

89 Reference pump cycle time.

90 Low oxygen calibration factor, sensor #1.

91 Low oxygen calibration factor, sensor #2.

92 Gas calibrate / check selection.

93 Oxygen mid point calibration factor #1.

94 Oxygen mid point calibration factor #2.

95 4-20mA output limit enable option.

96 Solid state relay fail alarm setting.

97 Heater solid state relay swapping function.


5.2

SET-UP & RUN MODES

For the SET-UP mode keyboard to operate, press the SET-UP/RUN button. The set-up
light will come on when the set-up mode has been entered.

NOTE:
Set-up mode cannot be entered if the keyboard lock switch (SW1) on the inside of the
transmitter is in the UP position. The keyboard lock switch can be found on the door PCB
(1630-1), on the lock side, at the top. If access is attempted while the keyboard is locked,
the message ‘Illegal Access’ will be displayed.

While the transmitter is in set-up mode the outputs will be frozen. All of the functions
written in BLUE will now operate. If there are not any buttons pressed for 1 minute the
transmitter will automatically revert to RUN mode.

If purges or an auto-calibration check occurs while the transmitter is in set-up mode, they
will be delayed until the transmitter is returned to RUN mode.
To cancel an automatic purge or calibration check cycle, press AUTO CAL button while in
RUN mode.
The Extended Menus can be seen if the Setup button is pressed before power is applied
to the transmitter, and held until the display shows the words “Extended Menu”.

5.3

FUNCTION SELECT

When the SET-UP mode is entered, the transmitter will automatically read the last set-up
function selected.

To select other functions, operate the ‘FUNCTION ’ button to increment to the next
function, or ‘FUNCTION ’ to decrement to the previous function.