Oxygen probe specifications, 9060-zirconia analyser – Teledyne 9060 - Zirconium oxide flue gas oxygen analyzer User Manual
Page 14
![background image](/manuals/345700/14/background.png)
9060-Zirconia
Analyser
OXYGEN PROBE SPECIFICATIONS
M
ODEL 1231
1232
Application
Combustion flue
Combustion flue gases
gases below
above 700°C (1290°F) with
900°C (1650°F)
no contaminants.
Refer to note 1
eg. natural gas, light oils
Temp. Range
0 to 900°C. Refer to note 2 700 to 1400°C
(32 to 1650°F)
(1470 to 2550°F)
Length
250 to 2000 mm
500 to 1500 mm
(10” to 80”)
(20” to 60”)
Process
1 ½" BSP
¾" BSP
Connection
or NPT
or NPT
Electrical
Weatherproof plug-in connector or optional screw terminals. The
Connection
plug connector can be supplied with the cable.
Cable
Order a specific length with the transmitter
Heater
Yes
No
Thermocouple
K, integral
R, integral
Response Time
Typically < 4 secs.
Typically < 1 sec
Head Temperature
-25 to 100°C (-15 to 210°F) with weatherproof connector
-25 to 150°C (-15 to 300°F) with screw terminals
Reference Gas
Ambient air 50 to 150 cc/min (3 to 9 scim). Pump can be supplied
with transmitter
Ref Air Connection
1/4" NPT
Integral air line in probe cable. Barbed
fitting to 3/16" ID PVC tube.
Filter
Removable sintered titanium alloy particulate filter,
30 micron standard, optional 15 micron available. Refer to note 2
CAL Check Gas
1/8" NPT female
1/8" NPT female
Connection
Weight
2 kg (4.4 lb) plus
1 kg (2.2 lb) plus
165 g (5.8 oz) / 100 mm
100 g (3.5 oz) / 100 mm
(4”) length
(4”) length