Teledyne 9060 - Zirconium oxide flue gas oxygen analyzer User Manual
Page 60
9060-Zirconia
Analyser
30. Transmitter Span Channel 2
Output
Zero Range
Span Range
Default Setting
SENSOR EMF
0 to 1100 mV in
100 to 1300 mV
0 to 100 mV
100 mV steps
in 100 mV steps
CARBON DIOXIDE
0 to 10 %
5 to 100 %
0 to 100 %
OXYGEN DEFICIENCY
+10 to -20 % oxygen 0 to 20% oxygen
-5 to +10 % oxygen
(see Note 3)
deficiency
excess
AUX TEMPERATURE
0 to 1300 °C
100 to 1400 °C
0 to 1300
°C
(32 to 2370°F)
(210 to 2550°F)
(32 to 2370°F)
in 100° steps
in 100° steps
LINEAR OXYGEN
0 to 99%
1 to100% oxygen
0 to 10% oxygen
LOG OXYGEN
0.1 % oxygen
20 % oxygen fixed
(see Note 1)
fixed
REDUCING OXYGEN
100% to 10 -10 % 10 -3 to 10 -30 %
100% to 10 -30 %
(see Note 2)
oxygen in one
oxygen in one
decade steps,
decade steps. Min
non
span five decades
overlapping
COMBUSTIBLES %
0 fixed
0.5 to 2.0 %
0 to 2.0 %
NOTE
1:
For log oxygen scale details, Refer to Appendix 3.
2:
Note that the reducing oxygen span is shown on the display as the exponent only. -1
represents
10
-1 % oxygen.
3:
The oxygen deficiency output can be used in the same way as a combustibles
analyser to signal the extent of reducing conditions. As an example, if the oxygen
deficiency is 3 %, then the burner would need 3 % oxygen
to bring it back to
stoichiometry.
4.
Always check that the “*” is on the RH end of the lower line to lock in the selection
before leaving the function.
5.
If output channel is selected for oxygen, and the temperature of sensor 1 or sensor 2
is below 650
°C (the display will be flashing “NOT READY”) the 4-20mA output for
the associated channel will be set to zero mA.
See set-up 91 for further details.