beautypg.com

Oxygen probe model selection guide, 9060-zirconia analyser – Teledyne 9060 - Zirconium oxide flue gas oxygen analyzer User Manual

Page 16

background image

9060-Zirconia

Analyser

OXYGEN PROBE MODEL SELECTION GUIDE


Heated probes-temperature range 0-900°C (1650°F).

1231 - U Length - Outer - Internal - Mounting

Sheath

Thermocouple

Thread

Basic model 2. 250mm (10”) 1. 316 SS max 850°C

1. Type K max 900°C

1. 1 ½ BSP

3. 500mm (20”) (1560°F)

(1650°F)

2. 1 ½ NPT

4. 750mm (30”) 2. Inconel *(1)

5. 1000mm (40”)

6. 1500mm (60”)

7. 2000mm (80”)


*Note:

(1) The Inconel option has all inconel wetted parts except for the ceramic sensor and viton ‘o’

rings.

Unheated probes for clean gases-temperature range 700-1400°C (1290-2550°F).

1232

-

U Length

-

Outer Sheath

-

Internal Thermocouple - Mounting Thread

Basic model 3. 500mm (20”)

1. 253 MA-max 1000°C

1. Nil *(2)

1. 3/4” BSP fixed

4. 750mm (30”)

(1830°F)

4. Type R max 1400°C 2. 3/4” NPT fixed

5. 1000mm (40”)

(2550°F)

6. 1500mm (60”)

3. High Purity Alumina

max 1300°C (2370°F) Horizontal

max 1400°C (2550°F) Vertical

4. 446 SS max 1000°C (1830°F)

*Note:

(2) A standard oxygen probe for carburising furnaces, has a 253 MA sheath.

1234 SENSOR SPECIFICATIONS

Range of measurement:

1 ppm to 100% oxygen

Output:

EMF = 2.154.10

-2

.T.log

e

(0.209/oxygen level of the sample)

Accuracy:

± 1%

Thermocouple: Type

K

Heater:

110 VAC 50 / 60Hz, 100 watts

Heater proportional band:

80

°C (175°F)

Speed of Response:

Less than 100 milliseconds

Sample flow rate:

1 to 5 litres / minute (120 to 600 scfm)

Differential Pressure:

80 to 800 mm (3 to 30”) WG gives a flow of 1 to 5 litres / min (120 to 600

scfm)
Process Connections:

1/4” NPT female, inlet and outlet

Dimensions:

300 mm (11.81”) high by 125 mm (4.92”) wide by 88 mm (3.46”) deep

Weight:

2.1 kg (4.6 lb)