Teledyne 9060 - Zirconium oxide flue gas oxygen analyzer User Manual

Page 55

background image

9060-Zirconia

Analyser

55

55

Options
Select the current year for the real time clock / calendar.

The cold start default sets the date and time to the software version date.

3. Calendar month

Options
Select the current month for the real time clock / calendar.

4. Calendar day

Options
Select the current day for the real time clock / calendar.

5. Real time8:14 PM clock hour

Options Select the current hour for the real time clock. (24 hour format)

6. Real time clock minutes

Options
Select the current minutes for the real time clock.

7. Reference voltage # 1 (about 27.5 mV)

Options
Set the value of the reference voltage as read on a 3 1/2 digit multimeter (See Section 6.2
for further details).
27.55 mV

*

8. Reference voltage # 2 (about 194 mV)

Options
Set the value of the reference voltage as read on a 3 1/2 digit multimeter (See Section 6.2
for further details).
193.60 mV

*

9. Reference voltage # 3 (about 1200 mV)

Options
Set the value of the reference voltage as read on a 3 1/2 digit multimeter (See Section 6.2
for further details).
1202.00 mV

*

10. Reference voltage # 4 (about 2500 mV)

Options
Set the value of the reference voltage as read on a 3 1/2 digit multimeter (See Section 6.2
for further details).
2479.00 mV

*


Set-up items 7 to 10 are used to calibrate the A/D of the instrument. This should be done
30 minutes or more after the instrument has been on, approximately once every year. The
calibration constants are retained in battery backed memory unless a ‘COLD START’ is
performed. Connect a 3 1/2 digit multimeter negative lead to the test point marked ‘C’ to
the right of the PCB on the inside of the door (labelled ‘REF VOLTS’). Measure the four
voltages on the test point marked 1 to 4 with the positive lead. Refer to Figure 6.2 in the