Recording with Roboocyte2 Control Software
77
7.5 Recording
Data
Manual Recording Mode
What You “Can Do”:
moving the measuring head into wells
measuring the resistance of the electrodes
determining electrode voltage offsets and compensate them
impaling oocytes, measure membrane potential, clamp either current or voltage
changing and testing various gain settings
(P gain, I gain, crosstalk capacitance compensation (CX))
switching on and off perfusion, switch between different solutions
recording and saving the data in continuous recording mode at various sampling frequencies
recording IV dependencies
What You “Can Not Do” (Roboocyte2 Version 1.3.1):
executing scripts
saving the recorded IV dependencies
linking solution exchanges to saved data from continuous recordings
7.5.1 Manual Recording Step by Step
1. Move the carrier to the respective well, for example to well A1. This will only work if you have
loaded or created a well plate before.
2. Move the measuring head down into the liquid (wells should be filled with at least 200 μl
of buffer).
3. Check the resistance of both glass microelectrodes. Resistances of a new measuring head should
be clearly below 1 MOhm.