Exporting a routine, Probe and heater tab – Luminex xPONENT 4.2 for FLEXMAP User Manual
Page 115

2. Click Import.
3. In the Open dialog box, browse for the file you want to import, then click Open. The
routine becomes the active routine.
Exporting a Routine
1. Open the Maintenance page, then open the Cmds & Routines tab.
2. Click Export.
3. In the Save As dialog box, navigate to the folder where you want to store the routine file,
then click Save.
NOTE: If you haven't previously saved a routine, the Export button will not
display.
Probe and Heater Tab
Maintenance > Probe & Heater
Use this tab to adjust the probe height and plate heater settings.
• Auto Adjust Height - Automatically adjusts the probe height using the specified location(s)
in the main plate, reservoir, and strip wells.
NOTE: Luminex recommends that you calibrate the probe height using well
D6 for a 96-well plate and well H12 for a 384-well plate.
Maintenance Page
101