Daily shutdown procedures – Luminex IS Version 2.3 (IVD) User Manual
Page 96

Luminex IS Software Manual for Version 2.3- For In Vitro Diagnostic Use
x
MAP Technology
86
the desired files to delete and click OK. Windows deletes the
files and closes the dialog box.
Figure 57 Disk Cleanup for (Selected Drive)
To delete the Message Log directory:
1. In the Cleanup Utility dialog box, click Delete MsgLog
Directory.
2. In the Cleanup Utility confirmation dialog box, click Yes.
To delete the batch folders:
1. In the Cleanup Utility dialog box, click Delete Batch
Directory.
2. In the Cleanup Utility confirmation box click Yes to delete all
the batch folders. All the folders under C:\My Batches are
deleted. The C:\My Batches folder remains.
Daily Shutdown
Procedures
Sanitize the
System
Sanitize the system with 10% to 20% household bleach to
decontaminate the sample lines and the cuvette after biohazard
contact. You should sanitize as part of your daily shutdown routine
after biohazard contact. Sanitizing uses the Luminex XYP reservoir
location because only the reservoir can accommodate the amount of
fluid necessary to sanitize the instrument.