Commands, Acquire patient, Add batch – Luminex IS Version 2.3 (IVD) User Manual
Page 39: Alcohol flush, Autoscale, Autosize, Backflush, Cal1, Cal2, Cancel (command)
x
MAP Technology
Luminex 2.3 Software
PN 89-00002-00-254 Rev. B
29
Commands
This section describes the different commands used in the Luminex
IS 2.3 software. They are grouped Alphabetically.
Acquire Patient
This command, which is part of the Insert command/menu, adds
patients to the command list. After adding the patients you define the
Sample ID and Dilution Factor for each new patient. The Dilution
Factor defaults to 1.
Add Batch
When creating a multibatch, adds an existing batch to a multi-batch.
Alcohol Flush
Removes air bubbles from the sample tubing and the cuvette using
70% isopropanol or 70% ethanol. The alcohol flush takes about five
minutes. Uses the Luminex XYP reservoir due to volume
requirement.
Autoscale
Automatically adjusts the maximum number of events shown on the
Y axis on the histogram. Click during acquisition to readjust the Y
axis scale.
Autosize
Automatically adjusts column widths to fit the data and header sizes.
Backflush
Removes obstructions from the fluidics pathways by drawing sheath
fluid from the sheath fluid container. You do not need to supply
solution in a plate. A backflush command takes about seven seconds.
CAL1
Runs the Classification portion of the Luminex System calibration.
CAL2
Runs the Reporter portion of the Luminex System calibration.
Cancel (Command)
Cancels the process for the last command initiated.
Cancel All
Cancels all the commands in progress.
Change Lot
Opens the Change Lot dialog box from which you select a lot to
apply to a batch.
CON1
Runs the verification for the Classification calibration.
CON2
Runs the verification for the Reporter calibration.
Connect to Instrument
Establishes a connection between the PC and the analyzer.