Of tab) table 5 lists the, Standards tab – Luminex IS Version 2.3 (IVD) User Manual
Page 33

x
MAP Technology
Luminex 2.3 Software
PN 89-00002-00-254 Rev. B
23
Figure 16 Analysis Window - Errors Tab Open
Standards Tab
The Standards tab lists all the batch tests with system comments
specific to each test. The detailed test information that is displayed
Table 5. Error Categories
System and Assay Errors
Sample Errors
•
Instrument Not Calibrated
•
Failed Verification (system lists
each failed control)
•
Failed Control (verifier failed)
•
Temperature Divergence from
Calibration Temperature
•
Failed Curve Fit
•
Analysis Error
•
APD Temp Range Exceeded
•
XYP Temp Unstable
•
Low Voltage
•
High Sheath Pressure
•
Low Sheath Pressure
•
Command Timeout
•
Low Laser Power
•
Cannot Calculate Inverse
Function
•
Failed Std in Batch
•
Insufficient Bead Count
•
Sample High/Low
•
Analyzer Error
•
High Sheath Pressure
•
Low Sheath Pressure
•
Sample Timeout
•
Sample Empty
•
Cannot Calculate Inverse
Function
•
Different Qualitative Results