beautypg.com

GF Signet 8850 ProcessPro Conductivity-Resistivity Transmitter User Manual

Page 3

background image

3

8850-3 Conductivity/Resistivity Transmitter

3.2 System Power/Loop Connections

Stand-alone application, no current loop used

Transmitter

Terminals

Power

Supply

12 to 24 VDC

Connection to a PLC with built-in power supply

Internal

PLC Connection

Connection to a PLC/Recorder, separate supply

Example: Two transmitters connected to PLC/Recorder with separate power supply

-

+

PLC or Recorder

Loop 2 -

Loop 2 +

System Power Loop 1 -

System Power Loop 1 +

AUX Power -

AUX Power +

6

5

4

3

2

1

Transmitter

Terminals

Loop 2 -

Loop 2 +

System Power Loop 1 -

System Power Loop 1 +

AUX Power -

AUX Power +

6

5

4

3

2

1

6

5

4

3

2

1

Transmitter

Terminals

Loop 2 -

Loop 2 +

System Power Loop 1 -

System Power Loop 1 +

AUX Power -

AUX Power +

Loop 2 -

Loop 2 +

System Power Loop 1 -

System Power Loop 1 +

AUX Power -

AUX Power +

6

5

4

3

2

1

6

5

4

3

2

1

Loop 2 -

Loop 2 +

System Power Loop 1 -

System Power Loop 1 +

AUX Power -

AUX Power +

Power

Supply

12 to 24 VDC

-

+

PLC or Recorder

Channel 3
4 to 20 mA in

Channel 4
4 to 20 mA in

-
+

-

+

Transmitter 1

Terminals

Transmitter 2

Terminals

PLC Terminals

Channel 1

4 to 20 mA

+
-

Power

Supply

12 to 24 VDC

Channel 2

4 to 20 mA

+
-

-
+

Channel 2

4 to 20 mA

+
-

Channel 1

4 to 20 mA

-
+

Channel 1
4 to 20 mA in

Channel 2
4 to 20 mA in

+
-

+

-

Power

Supply

12 to 24 VDC

-

+

NC

NC

AUX power required for all 8850 systems

1mS
1M

Ω

10mS

100K

Ω

1,000mS

10,000mS

50,000mS

200,000mS

100mS

10K

Ω

400,000mS

100,000mS

0.054mS

18.3 M

Ω

Bottled

Water

Brackish Sea Water

(Salinity)

UPW @ 25°C

USP

Cooling Tower

Waste Waters

Deionization Regen. Chemicals

Alkali Cleaners, Acids/Bases

Rinse

Water

3-2819, 3-2839

(0.01 Cell)

3-2820, 3-2840

( 0.10 Cell)

3-2821, 3-2841

(1.0 Cell)

3-2822, 3-2842

(10.0 Cell)

3-2823

(20.0 Cell)

0.010mS

100 M

Ω

200mS

Select the sensor with the range that best straddles your
application requirements.
• 2818-1, 2839-1 (0.01 cell):

0.01 to 100 μS/cm

(10

kΩ to 100 MΩ)

Use the 2818-1 or the 2839-1 for all resistivity requirements

• 2820-1, 2840-1 (0.1 cell):

1 to 1000 μS/cm

• 2821-1, 2841-1 (1.0 cell):

10 to 10,000 μS/cm

• 2822-1, 2842-1 (10.0 cell):

100 to 200,000 μS/cm

• 2823-1 (20.0 cell):

200 to 400,000 μS/cm